Recent Posts
Choosing a gas transmission rate tester for barrier films starts with a precise question: which gas must be measured, and under what conditions? ASTM D1434 addresses gas permeability of plastic film and sheeting using a pressure-based method. ASTM D3985 addresses oxygen gas transmission through plastic film and sheeting using a coulometric sensor. Their results can both inform barrier evaluations, but the measurement principles and reported quantities are not automatically interchangeable.
For technical evaluators, the decision affects more than instrument selection. It determines how a specimen is exposed, what the detector actually responds to, and whether a result can be compared with a customer specification or another laboratory’s data. The authoritative requirements are in the applicable editions of the standards; the comparison below is a selection aid, not a substitute for those documents. [Sources: ASTM D1434, https://www.astm.org/d1434.html; ASTM D3985, https://www.astm.org/d3985.html]
ASTM D1434 describes a differential-pressure approach. A film specimen separates a high-pressure side containing the test gas from a low-pressure side. Gas permeates through the film, and the method uses changes associated with the low-pressure side to determine permeation behavior. The standard describes both a manometric procedure and a volumetric procedure. [Source: ASTM D1434, https://www.astm.org/d1434.html]
Its practical appeal is that it is not inherently limited to oxygen. Depending on the apparatus, materials, safety controls, and the applicable procedure, a laboratory may evaluate films against different gases. That flexibility should not be confused with universal gas compatibility: seals, pressure components, gas handling, and instrument calibration must all be appropriate for the selected gas.
ASTM D3985 is specific to oxygen transmission. Oxygen on one side of the specimen permeates through the film and is transported to a coulometric sensor by a carrier-gas stream. The sensor response provides the basis for determining the oxygen gas transmission rate, often referred to as OTR. [Source: ASTM D3985, https://www.astm.org/d3985.html]
This is a useful fit when oxygen ingress is the central performance question—for example, when comparing candidate structures against an oxygen-barrier specification. A coulometric oxygen sensor does not make the instrument a general-purpose detector for nitrogen, carbon dioxide, or other gases. For those applications, assess a suitable method and detector separately.
Neither standard is a blanket certification of a film or an instrument. A report can state that a test was performed according to a specified standard and edition, subject to the actual procedure followed. Accreditation, where required, is a separate question: verify that the laboratory’s scope covers the method, material, and relevant measurement range. ISO/IEC 17025 is the general standard for testing and calibration laboratory competence. [Source: ISO, https://www.iso.org/standard/66912.html]
The appropriate method is often determined before a laboratory sees the sample. A customer drawing, material specification, regulatory submission, or internal qualification plan may name a standard, edition, conditioning regime, and reporting unit. If it does, establish those requirements before comparing testers.
For a technical evaluation, request a demonstration using a representative film where possible. Review the raw measurement trend, leak-check procedure, stabilization criteria, and final report—not only the number on the screen.
Transmission rate describes the amount of gas passing through a defined area over time under stated conditions. Permeability is a material-related quantity derived by accounting for factors including specimen thickness and the driving pressure difference, within the assumptions of the applicable method. The exact definitions and calculations must follow the standard and reporting convention used. [Source: ASTM D1434, https://www.astm.org/d1434.html]
That distinction matters when comparing films of different thicknesses. A thicker specimen may show a lower transmission rate, but dividing or multiplying by thickness does not automatically make every multilayer structure equivalent. Coatings, layers, defects, and processing history can influence performance; a simple thickness-normalized value should not replace a measurement on the actual structure being qualified.
Also check the pressure basis and units. Before comparing two reports, establish whether each value is a transmission rate or a permeability value, what area and time units are used, and how gas volume is referenced. Do not convert one result into another by applying a thickness factor alone.
A discrepancy is not, by itself, proof that one instrument is wrong. The methods use different measurement principles, and the tests may also differ in gas exposure, pressure, temperature, humidity, specimen preparation, or stabilization. Their outputs are not automatically equivalent simply because both concern oxygen barrier performance.
Investigate the comparison in this order: confirm the specimen identity and test conditions; check units and calculations; review mounting and leak checks; then examine calibration, blanks, repeatability, and whether each result lies within the instrument’s validated range. If a customer requires a particular method, resolve discrepancies through an agreed correlation study rather than substituting the other method unilaterally.
For very low-transmission films, a small mounting leak or a damaged test area can be significant relative to the gas passing through the specimen. A careful evaluation therefore extends beyond the sensor or pressure transducer to the entire test cell and workflow.
A technically useful quotation should connect the proposed configuration to your test requirement. Ask for written answers to the following, along with supporting documentation:
Paratronix Instruments provides packaging-testing instruments and testing solutions. When discussing a requirement with any supplier, share the governing specification and representative sample details rather than asking only for a tester “for high-barrier film.” The resulting technical discussion should make clear what the proposed configuration can and cannot support.
Not automatically. If the specification names ASTM D3985, obtain the customer’s agreement before using another method. A documented comparison may support an alternative for a particular material and purpose, but it does not make the two methods universally interchangeable.
No. Consider the validated range and uncertainty at your expected result, as well as leak control, conditioning, test duration, calibration, and repeatability. A low displayed number is not meaningful if the complete measurement system cannot support it reliably.
No. A film test characterizes the tested specimen under stated conditions. Finished-package performance can also be affected by seals, closures, folds, damage, and package geometry. If the decision concerns the completed pack, determine whether an appropriate package-level test is also required.
Leave A Reply
Search by Keywords


